Maintenance
Replacing the Pump Heads
5
Remove the complete pump head assembly by holding it
with both hands and pulling it to the front.
N O T E
Do not further disassemble the pump head.
7
Install flow connections.
Connect the degassing unit outlet to the inlet of the
primary pump head and the outlet of the secondary
pump head to the inlet of the purge valve.
Channel A (left pump head assembly) is connected to
port 4, channel B (right pump head assembly) to port 1.
138
6
Install the pump head assembly by fixing the 4 screws
using a 4 mm hex key and a torque wrench set to 5 Nm.
Install screws step by step, not screw by screw.
5 Nm
8
After the complete reassembly of the module, the pump
must be purged. For 10 min, apply a flow of 10 ml/min
using solvents suitable for your application and a
composition 50:50 for channels A and B, for example
50 % acetonitrile, 50 % water.