Introduction - Agilent Technologies Infinity Binary Pump VL 1290 User Manual

Table of Contents

Advertisement

8
Test Functions and Calibrations

Introduction

Introduction

All tests are described based on the Agilent Lab Advisor Software B.01.04.
Other user interfaces may not provide any test or just a few.
Table 8
Interface
Agilent Instrument Utilities
Agilent Lab Advisor
Agilent ChemStation
Agilent Instant Pilot
For details on the use of the interface refer to the interface documentation.
102
Interfaces and available test functions
Comment
Maintenance tests available
All tests are available
No tests available
Adding of pressure to
chromatographic signals
possible
Some tests are available
Available Function
• System pressure test
• System pressure test
• Pump head leak test
• Pressure
• Pressure ripple
• Temperature main board
• System pressure test
• Monitoring of values
Agilent 1290 Infinity Binary Pump User Manual
Pressure
Pressure ripple
Flow (in case of
operating pressure)