Troubleshooting and Diagnostics
Lab Advisor Service & Diagnostic
Pump Leak Rate Test for the
The test determines the leak rates in the primary and the secondary pump
chambers for component level diagnostic.
This 2D-LC test works only for the driver-based 2D-LC solution.
When
Diagnostic of the
Software required
Agilent Lab Advisor Software (2.17 or higher)
Preparations
Read the following:
•
•
Procedure to follow:
•
•
1 Select Service & Diagnostic from the menu.
2 Select the
3 Select the Pump Leak Rate Test.
4 Press the Run button and follow the instructions in the software.
2D-LC User Guide
2
D pump.
Documentation provided with the Agilent Lab Advisor online help
2D-LC Manual
Close the current Acquisition client window
Close instrument connection from OpenLab Control Panel
2
D pump.
Figure 151 Pump Leak Rate Test for the 2D-LC 1290 binary pump
2
D Pump