7
Method Development of Active Solvent Modulation (ASM)

Optimize the Sample Loop Flush

Optimize the Sample Loop Flush

Activate ASM in the software and set Flush sample loop to 3.0 times.
Flushing the sample loop 3 times is typically enough and the recommended
NOTE
default. Less time may be sufficient and can be verified during optimization. The
user interface displays how long this will take.
2D-LC User Guide
189