2D-LC Capillaries Configuration Tool

The Configuration tool of Agilent Lab Advisor stores by default only standard
capillaries. To add 2D-LC specific capillaries (e.g. Sample Loop, transfer capillary,
or ASM capillary) to the 2D-LC instrument, it is necessary to configure these
capillaries.
When
Installation of 2D-LC specific capillaries
Parts required
Description
All required capillaries for the 2D-LC setup
Software required
Agilent Lab Advisor Software (2.17 or higher)
Preparations
Read the following:
Take care that all capillaries are installed and their specification is available.
1 In Agilent Lab Advisor, start the Configuration tool.
2 Enter the specific parameters Length [mm] and Diameter [mm] for the Generic
3 Click send.
2D-LC User Guide
Documentation provided with the Agilent Lab Advisor online help
2D-LC Manual
The Edit Generic Capillaries function is available.
Sample Loop, Generic Transfer Capillary, and Generic ASM Capillary to the
fields.
Figure 145 2D-LC generic capillaries configuration
The Volume [µL] of the specified capillaries is automatically calculated.
The Configuration tool sends the parameters to the 2D-LC system.
The capillaries now appear in the Modify capillaries selection list of the
chromatographic data system.
283