11
Error Information
Module-Specific Error Messages
External Valve falls into resident mode
Error ID: Flashing status indicator
The valve drive was not able to operate correctly
Probable cause
1
2
2D-LC User Guide
Communication issues
Configuration mismatch
Suggested actions
•
Check the CAN cable connections of the
modules.
•
Check if the hosted module is present.
•
Check if the firmware on the entire stack is
out of the same firmware set.
•
Check if the limit of 3 hosted modules for
each host module is not exceeded.
•
Check if the dipswitch settings are correct.
•
Check if the firmware on the entire stack has
to be the latest version.
316