Error Information

General Error Messages

General Error Messages

General error messages are generic to all Agilent series HPLC modules and may
show up on other modules as well.

Timeout

Error ID: 0062
The timeout threshold was exceeded.
Probable cause
1
2
2D-LC User Guide
The analysis was completed successfully,
and the timeout function switched off the
module as requested.
A not-ready condition was present during a
sequence or multiple-injection run for a
period longer than the timeout threshold.
Suggested actions
Check the logbook for the occurrence and source
of a not-ready condition. Restart the analysis
where required.
Check the logbook for the occurrence and source
of a not-ready condition. Restart the analysis
where required.