Troubleshooting and Diagnostics
Lab Advisor Service & Diagnostic
Lab Advisor Service & Diagnostic

Decluster the 2D-LC Cluster

This tool allows to remove an LC device's clustering configuration data, e.g. the
linking between
When
Replacement of one of the cluster partners.
Software required
Agilent Lab Advisor Software (2.17 or higher)
Preparations
Read the following:
Procedure to follow:
1 Select Service & Diagnostic from the menu.
2 Select the
3 Select Firmware Declustering.
4 To Clear clustering configuration data, press the Run button.
To re-establish the link between the two modules, re-perform an Auto
NOTE
configuration and a selection as cluster.
2D-LC User Guide
2
D pump and 2D-LC valve.
Documentation provided with the Agilent Lab Advisor online help
2D-LC Manual
Close the current Acquisition client window
Close instrument connection from OpenLab Control Panel
2
D pump.
Figure 149 Firmware Declustering of the 2D-LC 1290 binary pump