Pump Head Leak Test For The 2D Pump - Agilent Technologies 1290 Infinity II 2D-LC User Manual

Solution for masshunter tof and q-tof
Hide thumbs Also See for 1290 Infinity II 2D-LC:
Table of Contents

Advertisement

10
Troubleshooting and Diagnostics
Lab Advisor Service & Diagnostic
Pump Head Leak Test for the
The test determines the leakage of the individual pump heads.
This 2D-LC test works only for the driver-based 2D-LC solution.
When
Diagnostic of the
Software required
Agilent Lab Advisor Software (2.17 or higher)
Preparations
Read the following:
Procedure to follow:
1 Select Service & Diagnostic from the menu.
2 Select the
3 Select the Pump Head Leak Test.
4 Press the Run button and follow the instructions in the software.
2D-LC User Guide
2
D pump.
Documentation provided with the Agilent Lab Advisor online help
2D-LC Manual
Close the current Acquisition client window
Close instrument connection from OpenLab Control Panel
2
D pump.
Figure 150 Pump Head Leak Test for the 2D-LC 1290 binary pump
2
D Pump
287

Advertisement

Table of Contents
loading

This manual is also suitable for:

1290 infinity ii lc

Table of Contents