Run The Experiment; Run The Checkout Familiarization Procedure For Standard Heart-Cutting 2D-Lc (Lc-Lc) - Agilent Technologies 1290 Infinity II 2D-LC User Manual

Solution for masshunter tof and q-tof
Hide thumbs Also See for 1290 Infinity II 2D-LC:
Table of Contents

Advertisement

8
Run the System

Run the Experiment

Run the Experiment

Run the Checkout Familiarization Procedure for
Standard Heart-Cutting 2D-LC (LC-LC)
To run the checkout, various hardware configurations are possible, see
on page 56. Not all options can be shown. As example the
is used here.
The following parameters have been optimized for this standard configuration.
Parameters can deviate slightly for your system. Run the experiment and
subsequently edit or optimize the method for your setup.
Table 15
Parameter
Column
Column temperature
Stop time
Mobile Phase A
Mobile Phase B
Flow Rate
Post time
Mobile Phase
Gradient:
Injection Volume
Injection Needle Wash In Flush Port, 10 s, acetonitrile/water (50/50)
Stop time
Diode-array Detector
Signal A
Reference Wavelength 360 nm
Reference Bandwidth
Peak width
Stop time
2D-LC User Guide
Recommended conditions in 1D (HPLC) for SHC, MHC, and HiRes 2D-LC
Value
1
D Column Compartment (MCT)
RRHD SB-C18, 2.1x 100 mm, 1.8 µm, 1200 bar (858700-902)
40 °C
As pump/No limit
1
D Pump
Water + 0.2 % formic acid
Methanol
0.6 mL/min
6 min
20 % B 0.00 min
100 % B 50 min
Autosampler
2 µLfor Standard LC
0.5 µL Positive Mode for LCMS
10 µL Negative Mode for LCMS
As pump/No limit
1
D Detector (DAD)
254 nm, Bandwidth 4 nm
100 nm
20 Hz
Stop time As pump/No limit
Table 5
Table 14
on page 198
201

Advertisement

Table of Contents
loading

This manual is also suitable for:

1290 infinity ii lc

Table of Contents