System Pressure Test for the
The test determines the leak tightness of the system between pump and blank
nut.
This 2D-LC test works only for the driver-based 2D-LC solution.
When
Leaks in the system flow path.
Tools required
Description
Wrench, 1/4 - 1/5 inch
Parts required
Description
Blank nut
Software required
Agilent Lab Advisor Software (2.17 or higher)
Preparations
Read the following:
Procedure to follow:
1 Select Service & Diagnostic from the menu.
2 Select the
2D-LC User Guide
Documentation provided with the Agilent Lab Advisor online help
2D-LC Manual
Close the current Acquisition client window
Close instrument connection from OpenLab Control Panel
2
D pump.
2
D Pump
289