Method Parameters
Set the 2D-LC Method parameters

Define the Posttime

To allow your column to equilibrate after changes in solvent composition (for
example after gradient elution), use the post time.
The instrument remains in a post-run state during the post time to delay the start
of the next analysis.
1 Check the Posttime radio button.
2 Specify the post time in the entry field.
2D-LC User Guide
The entry field becomes editable.
Limits: 0.01 – 99999 min.
Figure 72
Stoptime and Posttime settings
132