Recommended Tests To Conclude Troubleshooting - Agilent Technologies 1290 Infinity II 2D-LC User Manual

Solution for masshunter tof and q-tof
Hide thumbs Also See for 1290 Infinity II 2D-LC:
Table of Contents

Advertisement

10
Troubleshooting and Diagnostics

Recommended Tests to Conclude Troubleshooting

Recommended Tests to Conclude
Troubleshooting
The following table shows the most important tests to conclude troubleshooting.
Table 36
Recommended Tests for 2D-LC System Troubleshooting
Pump
Column
Compartment
Pressure Test
Thermostat Test
Leak Test
Pressure Test (if
column valve is
present)
Pump characteristic
Pump Ripple (1260 Pump)
Tuning (1290 pump)
2D-LC User Guide
For further detailed information, see:
Maintenance information in the specific manual of each module.
Troubleshooting Guide poster 5994-0709EN.
Best Practice for Using an Agilent LC System 01200-90090.
For additional help, contact your local Agilent Technologies service
representative.
Autosampler
Pressure Test
Inject standards or
inject different
volumina or blanks
Valve
Detector
Switching valve
Lamp Intentity
position/Check
Test
pressure reading
Wavelength
Pressure Test
calibration
In addition there
are detector
specific tests.
2D-LC Instrument
Run Checkout
For 2D-LC
Instruments
Pressure test
of the 1D-LC
Part
Pressure Test
of the 2D-LC
Part