2D-Lc Capillary Leak Test - Agilent Technologies 1290 Infinity II 2D-LC User Manual

Solution for masshunter tof and q-tof
Hide thumbs Also See for 1290 Infinity II 2D-LC:
Table of Contents

Advertisement

10
Troubleshooting and Diagnostics
Lab Advisor Service & Diagnostic

2D-LC Capillary Leak Test

Leak and tightness check of the 2D-LC Valve with the
the second dimension.
This 2D-LC test works only for the driver-based 2D-LC solution.
When
Leak in the 2D-LC valve.
Tools required
Description
Wrench, 1/4 - 1/5 inch
Parts required
Description
Blank nut
Software required
Agilent Lab Advisor Software (2.18 or higher)
Preparations
Read the following:
Procedure to follow:
1 Select Service & Diagnostic from the menu.
2 Select the
3 Select the 2D-LC Capillary Leak Test.
4 Press the Run button and follow the instructions in the software.
2D-LC User Guide
Documentation provided with the Agilent Lab Advisor online help
2D-LC Manual
Close the current Acquisition client window
Close instrument connection from OpenLab Control Panel
2
D pump.
2
D pump in the flow path of