Pressure Cluster Partner Missing; Position Cluster Partner Missing - Agilent Technologies 1290 Infinity II 2D-LC User Manual

Solution for masshunter tof and q-tof
Hide thumbs Also See for 1290 Infinity II 2D-LC:
Table of Contents

Advertisement

11
Error Information
Module-Specific Error Messages

Pressure Cluster Partner Missing

The connection from the valve drive to a defined pressure cluster partner is lost.
Probable cause
1
2

Position Cluster Partner Missing

Probable cause
1
2
2D-LC User Guide
Communication issues
Configuration mismatch
Communication issues
Configuration mismatch
Suggested actions
Check the CAN cable connections of the
modules.
Check and correct if necessary the valve
configuration and presence of defined pressure
cluster partner.
Suggested actions
Check the CAN cable connections of the
modules.
Check and correct if necessary the valve
configuration and presence of defined position
cluster partner.