Setting SatSeconds Alarm Limit
N-600x Operator's Manual
6. Wait for the display to time-out to accept the changes or press the
EXIT softkey to close the display and return to the normal
monitoring mode.
Note: Limit changes are in effect as long as the monitor remains
turned on. When the monitor is turned off, the institutional default
limits are restored. When the monitor is turned on, the institutional
default limits are effected. Factory or institutional defaults are
selected by qualified service personnel by following the procedures
outlined in the N-600x Service Manual.
Refer to Using SatSeconds on page 163, for more information on the
SatSeconds feature.
Note: The ability to adjust the SatSeconds Alarm limit can be enabled
or disabled by a qualified service personnel as described in the
N-600x Service Manual.
1. With the monitor in the normal monitoring mode, press the
LIMITS softkey. The current alarm limits display.
2. Press the SELECT softkey twice to select %SpO
3. Use the ADJUST UP or ADJUST DOWN buttons to select the
limit. The choices are 10, 25, 50, 100 seconds or OFF.
4. Press the EXIT softkey to save your selection.
Using the Monitor
SAT-S.