Using the Monitor
Setting Monitor Response Mode
66
The purpose of the response mode is to set the response time of the
N-600x algorithm calculation of the SpO
not affect the N-600x algorithm's calculation of pulse rate). The
trending interval (2- or 4-seconds) is updated automatically by the
monitor to roughly correspond with the SpO
time.
The response mode automatically programs the N-600x algorithm to
record and display monitor trend information at 2-second intervals
(Fast Mode) up to 24 hours or 4-second intervals (Normal Mode) up
to 48 hours.
The response mode display screen includes the current SpO
mode setting and the current measured %SpO
in the fast mode, the screen displays the fast mode symbol.
1. With the monitor in the normal monitoring mode, press the
SETUP softkey.
2. Press the NEXT softkey three times.
(the response mode does
2
calculation response
2
response
2
and pulse rate. When
2
N-600x Operator's Manual