Setting Up the Monitor
22
WARNING: Do not lift the pulse oximeter by the pulse oximetry
cable or power cord because the cable or cord could disconnect
from the pulse oximeter, causing the pulse oximeter to drop on
the patient.
WARNING: The N-600x is not defibrillator-proof. However, it
may remain attached to the patient during defibrillation or while
an electrosurgical unit is in use, but the readings may be
inaccurate during the defibrillation and shortly thereafter.
WARNING: In the USA, do not connect the pulse oximeter to an
electrical outlet controlled by a wall switch, because the pulse
oximeter may be accidentally turned off.
WARNING: Use only the Nellcor pulse oximetry cable DOC-10
with the N-600x pulse oximeter. Use of another pulse oximetry
cable will have an adverse effect on performance. Do not attach
any cable that is intended for computer use to the O
port. Do not connect any device other than a Nellcor-approved
O
M
sensor to the O
XI
AX
WARNING: The N-600x should not be used adjacent to or
stacked with other equipment. If adjacent or stacked use is
necessary, the N-600x should be observed to verify normal
operation in the configuration it is to be used.
Note: The monitor incorporates watchdog timers which reset the
monitor in the event of software errors.