Nellcor OxiMax N-600X Operator's Manual page 165

Hide thumbs Also See for OxiMax N-600X:
Table of Contents

Advertisement

Table 14: Common Problems and Resolutions
Problem
Error Code: "EEE XXX" followed
by a number appears.
N-600x Operator's Manual
Resolution
Press the ON/STANDBY button and allow the
monitor to shut off completely. Then press the
button again to turn the monitor back on.
If the error code persists, record the number and
provide this information to a qualified service
personnel, or your local Nellcor representative.
Error Code "EEE 529 or 729" displays when
the battery discharges to a critically low level.
Ensure the SUPPLY VOLTAGE SELECTOR
switch on the rear panel is set to the proper
voltage based on your location.
Press the ON/STANDBY button and allow the
monitor to shut off completely. Allow the
battery to charge for about 10 minutes and then
turn the unit back on.
If the error code is still present, turn the unit off
and allow it to continue to charge. If the
monitor has been charged for 30 minutes and
the error code persists, notify a qualified service
personnel, or your local Nellcor representative.
Troubleshooting
157

Advertisement

Table of Contents
loading

Table of Contents