Symbols, Controls, Displays and Indicators
Table 2: Visual Indicators
Indicator
Data In-Sensor
Indicator
SatSeconds
Indicator
Fast Response
Mode Indicator
Neonate Alarm
Limits Indicator
18
Description
Lights to indicate that the attached O
sensor event record. The sensor event record information may be
viewed or printed.
Fills in clockwise as the SatSeconds alarm management system
detects a %SpO
reading outside of the limit setting. Empties in
2
counterclockwise direction when %SpO
When the indicator is full, a medium priority alarm sounds.
Determines the response time (2 to 4 seconds in fast mode and 5 to
7 seconds in normal mode) applied by the O
calculation of SpO
. The O
2
rate is unaffected by the response mode setting. The trending
interval (2 seconds or 4 seconds) updated automatically by the
monitor to roughly correspond with the SpO
time.
Displays when the alarm limits are set to neonate. No symbol
displays when the monitor is set to adult limits.
M
sensor contains a patient
XI
AX
reading is within limits.
2
M
algorithm in its
XI
AX
M
algorithm's calculation of pulse
XI
AX
calculation response
2
N-600x Operator's Manual