Nellcor OxiMax N-600X Operator's Manual page 71

Hide thumbs Also See for OxiMax N-600X:
Table of Contents

Advertisement

Alarm Limit Changed Indicator
Setting Alarm Limits
N-600x Operator's Manual
Alarm limits that have been changed from the institution or factory
default settings are identified by a decimal point (.) after the displayed
reading (%SpO
or BPM). The changed parameter is also identified
2
by a decimal point on the alarm limits screen.
The Alarm Limit display enables you to adjust the upper and lower
saturation and pulse rate limits. It also enables you to adjust the
SatSeconds limit.
The Alarm Limit display is accessed by pressing the LIMITS softkey
on the main menu.
The Alarm Limit display includes the alarm limit table and current
measured %SpO
and pulse rate. The title of the alarm limit table
2
indicates whether the instrument is in Adult or Neonate monitoring
mode. If SatSeconds are enabled, the Alarm Limit display also
includes the SatSeconds indicator. Decimal points after the displayed
%SpO
or pulse rate indicate that the respective limits have been
2
changed from the power-on defaults.
Using the Monitor
63