Using the Data Port

Connecting to the Data Port

Data Port Pinouts

112
The N-600x data port may be connected to a serial printer or PC by
using a cable terminated with:
an AMP connector (AMP part number 747538-1),
ferrule (AMP part number 1-747579-2), and
compatible pins (AMP part number 66570-2).
The cable should not exceed 25 feet (7.6 meters) in length. The
external ITE (Information Technology Equipment) device must be
certified to UL-1950 or IEC-60950. The cable used must have a
braided shield that provides 100% coverage, such as a Belden cable
(Belden part number 9609) or equivalent. The shield must have a
360-degree connection to the metal shell on the N-600x's DB-15
connector and to the connector on the PC or serial printer.
Caution: Do not create sharp bends in the cable, as this may tear
or break the shielding.
No hardware flow control is used. However, in ASCII mode, XON/
XOFF flow control is supported.
The pinouts for the data port are listed in Table 8.
Table 8: Data Port Pinouts
Pin
Signal Name
1
-
RXD+ (RS
422 [+] input)
2
RXD_232 (RS
3
-
TXD_(RS
232 output)
4
-
TXD+ (RS
422 [+] output)
-
232 input)