Oximax Sensor Tabular Event Data - Nellcor OxiMax N-600X Operator's Manual

Hide thumbs Also See for OxiMax N-600X:
Table of Contents

Advertisement

Using OXIMAX Sensor Event Records
O
M
Sensor Tabular Event Data
XI
AX
96
The O
M
sensor tabular event data is a listing of all events
XI
AX
recorded on the O
M
XI
AX
SUMMARY
#
DATE
START
100
06JAN
13:55
06JAN
99
11:07
98
06JAN
10:30
97
06JAN
00:02
The table title is located in the upper left corner. Below the table title
is a six-column table with left-to-right column headings of event
number (#), date (DATE), event start time (START), event duration
(DUR), %SPO
minimum and maximum values during the event
2
(%SPO
), and pulse rate minimum and maximum values during the
2
event (BPM).
Event data are listed in chronological order with the most recent event
shown first, at the top of the list, when the tabular Event Summary
display is first accessed. Four events can be displayed
simultaneously; the table must be scrolled to view additional events.
You can move to the next screen view of the table, the next three
events (the previously displayed bottom or top event is retained as the
fourth event for context when a table is scrolled), using the two
left-most softkeys that are labeled with left- and right-facing arrow
icons, respectively. At the beginning of an event sequence, Event 1
of 5 events, the left-arrow softkey is blank; at the end of a sequence,
Event 5 of 5 events, the right-arrow soft key is blank, indicating the
beginning or end of the table.
sensor's memory chip.
DUR
%SPO2
00:03:00
75/80
00:10:30
76/83
00:06:30
79/84
00:02:00
82/84
BACK
PRINT
N-600x Operator's Manual
BPM
63/70
60/64
57/64
59/63

Advertisement