Trend Data in Graph Mode

Real-Time Display/Printout Format

N-600x Operator's Manual
Refer to Printing Monitor Trend Data on page 99 for the procedure to
print trend information. See Figure 4.
The graph mode disables all printout functions except trend data.
Graph mode trend printouts are formatted for a Seiko DPU-414 and
Okidata 320 serial printer.
Figure 4: Graph Mode Printout
Real-time data is continuously sent to the data port on the back of the
N-600x. Patient data can be obtained through the data port by
connecting the monitor data port to a PC or serial printer. When a
real-time printout or display is being transmitted to a printer or PC, a
new line of data is displayed every 2 seconds. Column headings are
displayed or printed after every 25 lines, or if one of the values in the
column heading changes. Readings are displayed at 4-second
intervals if the SpO
response mode is set to normal and at 2-second
2
intervals when the SpO2 response mode is set to fast.
Data cannot be obtained if the N-600x is operating on battery power.
Note: If the data output stops transmitting, turn the power off and
back on again or, if the monitor is connected to a PC, send an XON
(Ctrl-q) command to reset the monitor.