Nellcor OxiMax N-600X Operator's Manual page 192

Hide thumbs Also See for OxiMax N-600X:
Table of Contents

Advertisement

Specifications
184
Table 17: Electromagnetic Emissions
The N-600x is suitable for use in the specified electromagnetic
environment. The customer and/or user of the N-600x should assure that
it is used in an electromagnetic environment as described below:
Emissions Test
Compliance
RF emissions
Class B
CISPR 11
Harmonic
Complies
emissions
-
-
IEC 61000
3
2
Voltage
Complies
fluctuations/
flicker emission
-
-
IEC 61000
3
3
Table 18: Electromagnetic Immunity
-
The N
600x is suitable for use in the specified electromagnetic
environment. The customer and/or user of the N
is used in an electromagnetic environment as described below.
IEC
Immunity
-
60601
1
Test
Test Level
Electrostatic
±6 kV contact
discharge
(ESD)
±8 kV air
IEC
-
-
61000
4
2
Note: UT is the AC mains voltage prior to application of the test level.
Electromagnetic
Environment Guidance
-
The N
600x is suitable for use in
all establishments.
-
600x should assure that it
Compliance
-
2
Level
±6 kV contact
±8 kV air
N-600x Operator's Manual
Electromagnetic
Environment
Guidance
Floor should be
wood, concrete, or
ceramic tile. If floors
are covered with
synthetic material,
the relative humidity
should be at least
30%.