Troubleshooting
146
Table 12: Error Codes
Error
Error
Code
Message
80
DEFAULTS
LOST
81
SETTINGS
LOST
82
CLOCK
SETTING
LOST
515, 518,
N-600x Boot
534, 535,
Version x.x.x.x
569
529, 729
LOW
BATTERY
Action
-
The current power
on default settings
have been lost and returned to factory
defaults. Qualified service personnel
N-600x Service Manual
can use the
-
restore the desired power
on default
settings.
The current settings (for example, alarm
limits, alarm and pulse beep volumes,
alarm silence duration) have been lost
-
and returned to power
on defaults. Turn
the monitor off and back on again. If it
is necessary to have settings different
-
from the power
on default settings, turn
the monitor off and back on again, and
reenter the desired settings.
The date and time settings have been
lost. Reenter the date and time.
Recharge or replace the battery.
The application software is missing or
corrupt. Notify a qualified service
personnel.
The battery is discharged to a critically
low level. The monitor will shutdown
after 10 seconds.
Verify the SUPPLY VOLTAGE
SELECTOR switch on the rear panel is
set to the proper voltage.
Connect the monitor to AC power and
turn back on. A warning message is
displayed and a low priority audible
alarm sounded. These must be
acknowledged (by pressing the alarm
silence key twice) before the monitor
can be used for patient monitoring.
N-600x Operator's Manual