About The Audible Indicators - Nellcor OxiMax N-600X Operator's Manual

Hide thumbs Also See for OxiMax N-600X:
Table of Contents

Advertisement

About the Audible Indicators

N-600x Operator's Manual
Table 3: Audible Indicators
Function
Alarm Silence
Reminder
Confirmation Tone
Invalid Button Press
Valid Button Press
High Priority Alarm
Medium Priority
Alarm
Symbols, Controls, Displays and Indicators
Description
Three beeps sound approximately every three
minutes when alarms are silenced with the
alarm silence duration set to OFF and the alarm
silence reminder function is enabled.
Three beeps sound to indicate default settings
have been saved or reset to factory defaults or
trend data has been deleted.
-
Short, low
pitched tone indicating a button has
been pressed that is inappropriate for the
current state of the monitor.
-
Short, medium
pitched tone indicating an
appropriate button has been pressed.
-
-
High
pitched, fast
pulsing tone indicating
-
-
loss
of
pulse.
Note: If a High Priority Alarm is not silenced
within 30 seconds by pressing the ALARM
SILENCE Key, the monitor increases the
urgency level of the audible alarm signal by
alternating a piezo tone with the primary alarm
tone. See Piezo Tone on page 20.
-
Medium
pitched, pulsing tone indicating an
SpO
or pulse rate limit violation.
2
Note: If a Medium Priority Alarm is not
silenced within 2 minutes by pressing the
ALARM SILENCE Key, the monitor increases
the urgency level of the audible alarm signal by
alternating a piezo tone with the primary alarm
tone. See Piezo Tone on page 20.