Clearing Trend Information
N-600x Operator's Manual
5. Press the AMP softkey. The pulse amplitude units (PAU) trend
data displays.
The PAU reading (12 : 20) indicates the pulse amplitude units (upper
and lower) at the cursor position (dashed line). The cursor moves
right or left using the ADJUST UP (right) and ADJUST DOWN (left)
buttons.
1. With the monitor in the normal monitoring mode, press the
TREND softkey.
2. Press the MONITR softkey.
3. Press the NEXT softkey.
4. Press the DELETE softkey.
Note: Press NO and then EXIT to close this function without deleting
the trend data.
5. Press the YES softkey.
All the trend data clears and the monitor sounds three beeps.
Using Monitor Trend Data