Using the Monitor
Setting the Date and Time
56
WARNING: The sensor extrapolates from the date and time
provided by the N-600x when recording the sensor event record
to the sensor. The accuracy of the date/time is the responsibility of
the N-600x. It is recommended that the N-600x user set the time/
date to the correct value before a sensor event record-enabled
sensor is connected, and that this date/time not be changed while
the sensor remains connected. Since a sensor with sensor event
record data can be transported from one monitor to another,
having discrepancies in the date/time between monitors and the
sensor event record data will affect the order the sensor event
record data appears. To eliminate this possible problem, all
monitors within an institution should be set to the same time.
1. With the monitor in the normal monitoring mode, press the
SETUP softkey.
2. Press the NEXT softkey.
3. Press the CLOCK softkey.
4. Press the SET softkey.