Specifications
186
Table 18: Electromagnetic Immunity
-
The N
600x is suitable for use in the specified electromagnetic
environment. The customer and/or user of the N
is used in an electromagnetic environment as described below.
IEC
Immunity
60601
-
1
Test
Test Level
Voltage dips,
<5 % U
T
short
interruptions
and voltage
variations on
(>95 % dip in
power supply
U
) for 0.5
T
cycle
40 % U
T
IEC
-
-
61000
4
11
(60 % dip in U
for 5 cycles
70 % U
T
(30 % dip in U
for 25 cycles
<5 % U
T
(95 % dip in U
for 5 seconds
Note: UT is the AC mains voltage prior to application of the test level.
-
600x should assure that it
Compliance
-
2
Level
<5 % U
T
(>95 % dip in
U
) for 0.5
T
cycle
40 % U
T
)
(60 % dip in U
)
T
T
for 5 cycles
70 % U
T
)
(30 % dip in U
)
T
T
for 25 cycles
<5 % U
T
)
(95 % dip in U
T
T)
for 5 seconds
N-600x Operator's Manual
Electromagnetic
Environment
Guidance
Mains power quality
should be that of a
typical commercial
and/or hospital
environment. While
-
using the N
600x
requires continued
operation during
power mains
interruption, it is
recommended that
-
the N
600x be
powered from an
uninterruptible
power supply or
battery.