Oximax Sensor Event Record Graphical Data - Nellcor OxiMax N-600X Operator's Manual

Hide thumbs Also See for OxiMax N-600X:
Table of Contents

Advertisement

Using OXIMAX Sensor Event Records
O
M
Sensor Event Record Graphical Data
XI
AX
92
Graphical representations of patient event history is only available on
single-patient-use O
M
XI
AX
minimum or maximum %SpO
throughout the duration of an event (%SpO
alarm threshold for at least 15 seconds) and continuing every 30
seconds until the actual %SpO
threshold.
The duration of an event is determined by the number of data points
in the event. Each data point is stored at 30-second intervals.
Events end due to one of the following reasons:
%SpO
returns to or above the alarm limit
2
Loss of pulse
O
M
sensor is disconnected
XI
AX
O
M
sensor is removed from the patient
XI
AX
The graph title shows the data type (EVENT GRAPH) in the upper
left corner. The number of the displayed event and the total number of
events recorded in the O
title (example, 2/2). The date and time of the displayed event are
shown in the upper center and upper right corner.
sensors. Graphed data points are the
value for each 30-second interval
2
continuously below
2
value equals or exceeds the alarm
2
M
sensor are shown to the right of the
XI
AX
N-600x Operator's Manual