Parameter Ranges - Nellcor OxiMax N-600X Operator's Manual

Hide thumbs Also See for OxiMax N-600X:
Table of Contents

Advertisement

Using the Monitor

Parameter Ranges

40
The parameters of the N-600x monitor are preset to factory default
settings. See Factory Default Settings on page 167. The factory
default parameters may be changed to institutional default parameters
by following the procedures in the N-600x Service Manual.
Table 6 lists the parameters, ranges available, and the factory default
setting. The parameters may be set on an individual basis, by the
clinician, and these settings remain in effect until the monitor is
turned off.
Table 6: Parameter Ranges
Ranges/
Parameter
Selections
%SpO
Lower Alarm
2
Upper Alarm
Limit plus 1 to
Limit
100%
%SpO
20% to Upper
2
Lower Alarm
Alarm Limit
Limit
minus 1
Pulse Rate
Lower Alarm
Upper Alarm
Limit plus 1 to
Limit
250 bpm
Pulse Rate
30 bpm to Upper
Lower Alarm
Alarm Limit
Limit
minus 1
Alarm
Alarms 30, 60,
Silence
90, 120 seconds
Duration
Alarm
1 to 10
Volume
Alarms
Allow Off
Yes/No
Off Reminder -
Yes/No
Factory
Adult Defaults
100%
85%
170 bpm
40 bpm
60
7
-
No
Yes
N-600x Operator's Manual
Factory
Neonate
Defaults
95%
85%
190 bpm
90 bpm
60
7
No
Yes