Table 4 Equipment Required To Test Digital Channel Threshold Accuracy - Agilent Technologies InfiniiVision 7000 Series Service Manual

Hide thumbs Also See for InfiniiVision 7000 Series:
Table of Contents

Advertisement

2
Testing Performance
30
Equipment Required to Test Digital Channel Threshold
Table 4
Accuracy
Equipment
Digital Multimeter
Oscilloscope Calibrator
BNC-Banana Cable
BNC Tee
50
BNC Cable
BNC Test Connector,
8-by-2
Probe Cable
1 Turn on the test equipment and the oscilloscope. Let
them warm up for 30 minutes before starting the test.
2 Set up the oscilloscope calibrator.
a Set the oscilloscope calibrator to provide a DC offset
voltage at the Channel 1 output.
b Use the multimeter to monitor the oscilloscope
calibrator DC output voltage.
3 Use the 8- by- 2 test connector and the BNC cable
assembly to connect digital channels D0- D7 to one side of
the BNC Tee. Then connect the D0- D7 ground lead to the
ground side of the 8- by- 2 connector. See
Critical
Recommended Model/Part
Specifications
0.1 mV resolution,
Agilent 34401A
0.005% accuracy
DC offset voltage
Fluke 5820A
6.3 V
Agilent 11001-66001 or
Pomona 2BC-BNC-36
Agilent 1250-0781 or
Pomona 3285
Agilent 10503A
User-built (See
Agilent 01650-61607 or
Agilent N6450-60001
7000 Series Oscilloscopes Service Guide
page
26)
Figure
2.