To Verify Time Scale Accuracy; Table 6 Equipment Required To Verify Time Scale Accuracy - Agilent Technologies 5000 Series Service Manual

Hide thumbs Also See for 5000 Series:
Table of Contents

Advertisement

To verify time scale accuracy

5000 Series Oscilloscopes Service Guide
Time Scale Accuracy is a warranted specification for 5000
Series oscilloscopes manufactured after January 1, 2008. |
If your oscilloscope's serial tag has old-style (linear) barcodes
instead of the 3-D barcodes shown above, or if the date
indicated on the serial tag is older than January 1, 2008, then
disregard this test.
This test verifies the accuracy of the time scale. In this test you
will measure the absolute error of the timebase oscillator and
compare the results to the specification.
Table 6
Equipment Required to Verify Time Scale Accuracy
Equipment
Critical Specifications
Signal Generator
100 kHz - 1 GHz, 0.01 Hz
frequency resolution,
jitter: < 2ps
Cable
BNC, 3 feet
1 Set up the signal generator.
a Set the output to 10 MHz, approximately 1 V
2 Connect the output of the signal generator to oscilloscope
channel 1 using the BNC cable.
Testing Performance
Date Code:
0747
07 = Year
47 = Week
Recommended
Model/Part
Agilent N5181A,
E4400B, or 8648A
Agilent 10503A
sine wave.
pp
2
33