Exchange The Metering Seal And Piston - Agilent Technologies InfinityLab LC Series User Manual

Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

9
Maintenance and Repair

Exchange the Metering Seal and Piston

Exchange the Metering Seal and Piston

When
Poor injection-volume reproducibility
Leaking metering device
Tools required
p/n
8710-0510
8710-2392
8710-2411
01018-23702
Parts required
p/n
0905-1503
5067-5678
0905-1717
5067-5920
Preparations
Finish any pending acquisition job and in order to avoid leaks, stop the pump running.
When the instrument setup has changed, configure the new setup in the online view of
N O T E
the Chromatographic Data System. See
212
Description
Open-end wrench 1/4 — 5/16 inch
4 mm Hex key
3 mm Hex key, 12 cm long
Insert tool
Cleaning tissue and appropriate solvent like isopropanol or methanol
Description
Metering seal, 100 µL
Piston ceramic
for 100 µL analytical head
Metering seal, 40 µL
Piston ceramic
for 40 µL analytical head
Table 12
Agilent InfinityLab LC Series Vialsamplers User Manual
on page 120.

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents