9
Maintenance and Repair
Exchange the Wash Port Assembly
18
Reconnect the sample-loop to the needle.
N O T E
Do not overtighten the fitting!
N O T E
Incorrect positioning and installation of the
needle/loop connection can result in damaging and
breaking the sample loop.
20
Turn or flap the wash port to the right.
226
19
In Lab Advisor use Next to lift the needle slowly into the
up position.
21
Install the safety cover (1.) and fix the screw (2.).
Agilent InfinityLab LC Series Vialsamplers User Manual