14
Reconnect the wash tubing fitting to the new wash port
assembly.
16
Install and reconnect the wash tubing to the peristaltic
pump.
Agilent InfinityLab LC Series Vialsamplers User Manual
Exchange the Wash Port Assembly
15
Reinstall the new wash port assembly by mounting the
flapper into the housing of the needle station. The pin
must be latched into the housing. Check the movement
of the flapper. Check if the tension on the washport
spring is large enough that the part well springs back
into the original position.
17
Insert the needle (1). Align the needle in the seat, then
tighten the screw firmly (2).
Maintenance and Repair
9
225