16
In Lab Advisor use Next to lift the needle slowly into the
up position.
18
Install the safety cover (1.) and fix the screw (2.).
21
Close the doors.
Agilent InfinityLab LC Series Vialsamplers User Manual
Exchange the Needle Seat Assembly
17
Turn or flap the wash port to the right.
19
Select Next to move the needle into the needle seat.
20
Select Back to finish the procedure.
22
Perform a pressure test.
Maintenance and Repair
9
201