Agilent Technologies InfinityLab LC Series User Manual page 149

Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

To prepare autosampler for transportation
The park arm function moves the gripper and transport slider to the home
position. Reinstall the transport foam to secure the instrument against a
mechanical stop. The autosampler can be switched OFF after the preparation.
When
Before transporting or shipping the module.
Unsecured Transportation of the module
C AU T I O N
Unsecured transportation of the module may result in mechanical damage to the
gripper and transport slider.
Before parking the gripper arm, ensure there is no vial in the gripper. Use the Release
N O T E
Gripper function to remove the vial.
1 Select Park Arm.
2 When the arm is in the park position, install Transport-Protection
Agilent InfinityLab LC Series Vialsamplers User Manual
Always secure the arm in the park position plus transport foam.
(G7129-40050). Then the autosampler is ready for shipment, and can be
switched OFF.
Troubleshooting and Diagnostics
Maintenance Functions
7
149