1
In the Local Controller start the maintenance mode and
select Change metering device function.
OR
In the Tools section of the Agilent Lab Advisor software
select Service & Diagnostics > Maintenance Positions
> Change Metering Device, click Start and wait until
the metering device is in maintenance position.
2
Open the front door.
4
Remove the two fixing bolts.
6
Remove the two fixing bolts from the base of the
metering head assembly.
Agilent InfinityLab LC Series Vialsamplers User Manual
Exchange the Metering Seal and Piston
3
Disconnect the two capillaries from the metering device.
5
Pull the metering head assembly away from the sampler.
7
Disassemble the metering head assembly.
Maintenance and Repair
9
213