Agilent Technologies InfinityLab LC Series User Manual page 227

Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

22
Select Next to move the needle into the needle seat.
23
Select Back to finish the procedure.
25
Purge the wash port. Check if solvent will be delivered from the peristaltic pump.
Agilent InfinityLab LC Series Vialsamplers User Manual
Exchange the Wash Port Assembly
24
Close the doors.
Maintenance and Repair
9
227