Exchange the Wash Port Assembly
8
Disconnect the sample loop from the needle.
10
Lift out the needle.
12
Release the spring loaded pin.
224
9
Loosen the fixing screw.
11
Disconnect the wash tubing from the peristaltic pump.
13
Lift the flapper out of the bracket. Be careful not to
squeeze the wash tubing.