Maintenance and Repair
Exchange the Wash Port Assembly
8
Disconnect the sample loop from the needle.
10
Lift out the needle.
12
Release the spring loaded pin.
224
9
Loosen the fixing screw.
11
Disconnect the wash tubing from the peristaltic pump.
13
Lift the flapper out of the bracket. Be careful not to
squeeze the wash tubing.
Agilent InfinityLab LC Series Vialsamplers User Manual