Exchange the Sample Loop Assembly
6
Disconnect the loop capillary from the metering device
and from the needle.
8
After installing the new sample loop, check the needle
tip position by using the Needle Control function in Lab
Advisor.
10
Turn or flap the wash port to the right.
204
7
Install the new sample loop.
N O T E
Incorrect positioning and installation of the
needle/loop connection can result in damaging and
breaking the sample loop.
9
Click on Next in the LabAdvisor Needle lift up again.
11
Install the safety cover (1.) and fix the screw (2.).