Exchanging the Metering Seal
When
Poor injection-volume reproducibility
Leaking metering device
Tools required
p/n
8710-2411
Parts required
#
1
1
Preparations
• Select Start in the Tools function in the LMD software Change piston.
• Remove the upper front cover.
1
Remove the two capillaries from the metering-head
assembly.
284
Description
Wrench, 1/4 inch (supplied in HPLC Tool-Kit)
Hexagonal key, 4 mm (supplied in HPLC Tool-Kit)
Hexagonal key, 3 mm (supplied in HPLC Tool-Kit)
p/n
Description
5063-6589
Metering seal (pack of 2) for 100 µL analytical head
5063-6586
Sapphire piston
2
Remove the two fixing bolts, and pull the head assembly
away from the sampler. Notice that the closed side of
the metering head faces upwards.