Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

3
Install the capillary connection from the multisampler to
the Integrated Column Compartment.
5
Tighten the capillary fitting to make the connection
pressure tight.
N O T E
Do not overtighten the fitting (maximum torque
2.5 Nm).
N O T E
The use of capillaries with SL or SX fittings are
recommended for a better handling.
Agilent InfinityLab LC Series Vialsamplers User Manual
4
Connect the capillary to the heat exchanger.
6
Install the column holders to the heat exchanger.
Using the Module
Install a Column (ICC)
3
73