Agilent Technologies InfinityLab LC Series User Manual page 220

Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

9
Maintenance and Repair
Replace the Peristaltic Pump Cartridge
1
Open the front door.
3
Pull the cartridge forward off the motor shaft.
5
Push the new cartridge onto the motor shaft until the
clips click into place.
220
2
Press the two clips on the front of the peristaltic pump
cartridge.
4
Disconnect the tubing coupler leading to the wash port
and the tubing coupler coming from the solvent bottle.
6
Connect the wash port tubing to the upper tubing of the
new cartridge (use sand paper to get a good grip on the
tubing).