Transporting the sampler with a sample cooler or sample thermostat installed is only
N O T E
allowed for short distances. For longer distances, you must separate the units and send
them independently.
Mechanical damage of the module
C AU T I O N
If the transport assembly is not parked and not protected by the transport foam, the
module could be damaged due to excessive shock of the shipping container during
transport.
If the sampler with a sample cooler or sample thermostat needs to be shipped
to another location via carrier, ensure:
• The two modules are shipped in separate boxes.
• The gripper arm of the sampler is parked, see Park Arm in Agilent Lab
• Install the transport protection.
• Remove all vials from the drawers.
• The condensed water inside of the sample cooler or sample thermostat is
Agilent InfinityLab LC Series Vialsamplers User Manual
Transporting the Sampler with a Sample Cooler or Sample Thermostat Installed
Always park the transport assembly before shipment.
Protect the instrument by Transport-Protection (G7129-40050)
Store the installation foam in a save place, to use it for later transport of the
module.
Advisor online help for more information. Also ensure there is no vial in the
gripper arm.
removed.
3
Using the Module
85