4
Unscrew the safety cover screw.
WA R N I N G
Sharp needle
Uncovered needles may cause injuries.
➔
Do not change the needle seat at this point.
➔
Do not touch the tip of the needle.
6
Turn or flap the wash port to the left.
Agilent InfinityLab LC Series Vialsamplers User Manual
Exchange the Wash Port Assembly
WA R N I N G
Sharp needle
Uncovered needles may cause injuries
➔
Do not touch the tip of the needle.
5
Remove the safety cover.
7
In Lab Advisor use Next to move the needle in the down
position.
The needle will move slowly downwards and will stop
approximately 2 mm above the seat.
Maintenance and Repair
9
223