2
Site Requirements and Specifications
Site Requirements
Room Size and Ventilation
Flammable refrigerant
WA R N I N G
Formation of flammable gas-air mixtures inside the sample thermostat and
laboratory.
➔
➔
➔
➔
38
Keep open fire or sources of ignition away from the device.
Ensure a room size of 4 m
sample thermostat).
Ensure adequate ventilation: typical air exchange of 25 m
floor area.
Keep all ventilation openings in the enclosure clear of obstructions. Do not block
the openings on the circumference of the sample thermostat.
3
3
(1 m
for every 8 g of R600a refrigerant inside of the
Agilent InfinityLab LC Series Vialsamplers User Manual
3
2
/h per m
of laboratory