Agilent Technologies InfinityLab LC Series User Manual page 217

Vialsamplers
Hide thumbs Also See for InfinityLab LC Series:
Table of Contents

Advertisement

4
Rotate the arm approximately 2.5 cm (1 inch) to the left
and insert the straightened paper clip into the slit.
6
Hold the paper clip in place, press the gripper release
button and rotate the gripper arm to the right.
8
Replace the gripper arm by holding the paper clip in
place, pushing the gripper arm into the holder and
rotating the gripper arm to the left.
Agilent InfinityLab LC Series Vialsamplers User Manual
Exchange the Gripper Arm
5
Rotate the gripper arm slowly from left to right and apply
a gentle pressure to the paper clip.
The clip will engage on an internal catch and the rotation
of the arm will be blocked.
7
The gripper arm will come off.
9
In the Local Controller close Change gripper.
OR
In the Agilent Lab Advisor software select Change
gripper > Next, click Next and wait until the gripper is in
the default position.
10
Select Back to finish the procedure.
Maintenance and Repair
9
217